Product Specifications • Ambient temperature: 5 °C ~ 40 °C • Relative humidity: 20% ~ 90% • Range of atmospheric pressure: 700hPa ~ 1060hPa • The syringe pump shall be operated in the environment free from strong compact, vibration, water, other liquid penetrating into the unit, and corrosive gas. Transportation and storage conditions: • Ambient temperature: -40 °C ~ + 60 °C • Relative humidity: # 90% • Range of atmospheric pressure: 700hPa ~ 1060hPa • Overall dimension: 308mm (length) 144mm (width) 156mm (height) • Net weight : 2.2Kg
• Speed : 50 ml syringe : 0.1 ml/h ~ 1200 ml/h (0.1 ml/h each level) 30 ml syringe : 0.1 ml/h ~ 600 ml/h (0.1 ml/h each level) 20 ml syringe : 0.1 ml/h ~ 400 ml/h (0.1 ml/h each level) 10 ml syringe : 0.1 ml/h ~ 300 ml/h (0.1 ml/h each level) 5 ml syringe : 0.1 ml/h ~ 100 ml/h (0.1 ml/h each level)
• Speed accuracy: the error of infusion speed shall not exceed ± 3%
• Fast infusion: 1200 ml/h (50 ml syringe) 600 ml/h (30 ml syringe) 400 ml/h (20 ml syringe) 300 ml/h (10 ml syringe) 100 ml/h (5 ml syringe)
• Total amount inquiry: 0.1 ml ~ 9999.9 ml (0.1 ml each level)
• Limited amount: 0.1 ml ~ 9999.9 ml (0.1 ml each level)
• Alarm: Residual alarm, completed injection alarm, tubeblockage alarm, unclamped cylinder alarm, improperly installed putter alarm, out-of-limit speed alarm, alarm of infusion equivalent to limited amount, disconnected power line alarm, battery power depletion alarm, system error alarm, neglected operation alarm.
• Power supply: AC 100-240 V, 50/60 Hz; in-built batery DC 11.1 V rechargeable lithium batery; the fully charged batery can sustain the pump operaaonor more than 3 hours at the speed of 5 ml/h (specified by EN 60601-2-24:1998 standard).